Test Kit Chloride H/R 100-20 000ppm
Test Kit Chloride H/R (Drop Test)100-20 000ppm
The Chloride High Range (H/R) Drop Test Kit (100–20,000 ppm) is used for rapid, on-site, high-concentration chloride monitoring in industrial water, boiler, cooling tower, and wastewater systems. It uses a titration method (typically mercuric nitrate) to detect corrosive salinity.
Key Applications and Uses:
Industrial Water Systems: Monitors high chloride levels in boilers, cooling towers, and industrial processes to prevent corrosion.
Wastewater Analysis: Measures chloride concentrations in wastewater, agricultural runoff, and treated effluent.
Environmental Monitoring: Detects contamination from industrial discharges, mining, or road salt.
Process Water Quality: Ensures compliance in manufacturing processes requiring specific salinity levels. Test Kit Chloride H/R 100-20 000ppm
Test Kit Chloride H/R 100-20 000ppm
Test Kit Chloride H/R (Drop Test)100-20 000ppm
Key Features:
High Range: Capable of measuring up to 20,000 mg/L (ppm).
Methodology: Utilizes a drop-count titration method, often with Mercuric Nitrate, to provide quick, on-site results.
Kit Contents: Typically includes a calibrated,, reagent bottles, and a sample container.
Masiye Labs
Masiye Labs
+4
Operational Steps:
Sample Collection: Collect the water sample, sometimes requiring dilution for very high concentrations.
pH Adjustment: Use nitric acid or similar reagents to adjust the pH to approximately 3 (often turns yellow).
Titration: Add the titrant (e.g., Mercuric Nitrate) drop-by-drop while shaking until a color change occurs (typically from yellow to purple/violet).
Calculation: Count the drops and multiply by the factor provided in the instructions to determine the ppm.