Biological Safety Cabinet HEPA
Biological Safety Cabinet HEPA
Lead Time: 4-6 Weeks
A Class II Type A2 Biological Safety Cabinet (BSC) uses HEPA filters to create a sterile work zone, protecting personnel, samples, and the environment by filtering air to safeguard against biohazards like bacteria, viruses, and cell cultures, commonly used in microbiology, molecular biology, drug compounding, and diagnostics, but not for volatile chemicals or open flames. Biological Safety Cabinet HEPA Filter A2 | Biological Safety Cabinet HEPA
Key Uses & Applications
Biological Safety Cabinet HEPA Filter A2 | Biological Safety Cabinet HEPA
Microbiology & Virology: Safe handling of bacterial, viral, and microbial cultures (Biosafety Levels 1-3).
Cell Culture: Maintaining aseptic conditions for tissue culture, media preparation, and subculturing.
Pharmaceuticals: Sterile drug compounding, formulation, and quality control.
Clinical Diagnostics: Preparing PCR samples, handling patient specimens, and infectious disease testing.
Molecular Biology: DNA/RNA extraction and recombinant DNA work requiring contaminant-free environments.
How it Works (with HEPA Filter)
Inward Airflow: Creates a protective barrier at the front opening to keep contaminants out.
Vertical Downflow: HEPA-filtered air flows vertically down into the work zone, protecting samples.
Recirculation & Exhaust: About 70% of air is recirculated through the supply HEPA filter, while the remaining 30-40% is exhausted through a second HEPA filter, protecting the room and environment.
What NOT to Use It For
Volatile Toxic Chemicals: The exhaust air must not be recirculated into the lab if volatile chemicals are used; it must be ducted outside.
Flammable Materials: Open flames disrupt airflow and pose fire risks.
Highly Toxic Agents: For agents requiring total containment, a Class II Type B2 cabinet or fume hood might be needed.
The Special Cabinets are manufactured upon an order.
Biological Safety Cabinet HEPA Filter A2 | Biological Safety Cabinet HEPA
Biological Safety Cabinet Class 99.999% efficient HEPA Filter A2 | Biological Safety Cabinet HEPA
1. Small dimension,save space.
2. Centrifugal fan, speed, adjustable; H14 HEPA filter.
3. Large LCD diaplay: Operators can check detailed status of the cabinet, such as inflow and downflow velocity, work area temperature and humidity, filter pressure, UV working time and filter working time, filter life indicator etc.
4. Time reaerve function: This can save 30 minutes waiting time after activating the cabinet and the sterilization time after experiment.
5. Remote control: Each function can be realized 6 meters away from the cabinet by remote control, which can protect the operator under emergency.
6. Motorized front window: The front window is motorized for convenient, one-hand operation.
7. Voice prompt function: Airflow alarm, filter replacing alarm and front window height alarm.
Parameter of class ii biological safety cabinet
| Model | 11231BBC86 |
| External Size(W*D*H) | 700*650*1230 mm |
| Internal Size(W*D*H) | 600*500*540 mm |
| Tested Opening | Safety height: 200mm(8″) |
| Max Opening | 370 mm(15’’) |
| Inflow Velocity | 0.53±0.025 m/s |
| Down Flow Velocity | 0.33±0.025 m/s |
| HEPA Filter | Two, 99.999% efficient at 0.3μm. Filter life indicator. |
| Front Window | Motorized. Two-layer laminated toughened glass ≥ 5mm.
Anti UV. |
| Noise | EN12469 ≤ 58 dB / NSF49 ≤ 61 dB |
| UV Lamp | 15W*1 |
| UV timer, UV life indicator, emission of 253.7 nanometers
for most efficient decontamination. |
|
| Fluorescent Lamp | 40W*1 |
| Illumination | ≥1000Lux |
| Consumption | 500W |
| Socket | Two, total load of two sockets: 500W |
| Display | LCD display |
| Control System |
Microprocessor |
| Airflow System | 70% air recirculation, 30% air exhaust |
| Visual and Audio alarm | Filter replacement, window over height, abnormal air flow velocity. |
| Material | Work Zone: 304 stainless steel |
| Material | Main Body: Cold-rolled steel with anti-bacteria powder coating. |
| Work Surface Height | 750mm with optional base stand |
| Power Supply | AC220V±10%, 50/60Hz; 110V±10%,60Hz |
| Standard Accessory | Fluorescent lamp, UV lamp*2, Remote control,
Waterproof sockets*2 |
| Optional Accessory | Base stand, Universal caster with brake and leveling feet. |
| Gross Weight without Base Stand | Wooden box: 164kg Carton:121kg |
| Gross weight with Base Stand | Wooden:196kg |
| Package Size without Base Stand (W*D*H) | Wooden:860*800*1450mm Carton:820*760*1430mm |
| Package Size with Base Stand (W*D*H) | Wooden: 860*960*1450mm |
- Type: Laboratory Furniture
- General Use: Commercial Furniture
- Material: Metal
- Metal Type: Stainless Steel
- Place of Origin: Shandong, China
- Brand Name: BIOBASE
- Certification: ISO9001
- Accessory: Socket
- Color: White
- External Size(W*D*H): 700*650*1230 mm
- Tested Opening: Safety height: 200mm(8″)
- Inflow Velocity: 0.53±0.025 m/s
- Down Flow Velocity: 0.33±0.025 m/s
- Noise: EN12469 ≤ 58 dB / NSF49 ≤ 61 dB
Biological Safety Cabinet HEPA